The Agilent 66111A fast transient DC source is a low cost solution that delivers high throughput and exceptional sourcing capabilities. The 66111A is ideal for applications that do not require the dynamic and microamp measurement capabilities of a 66311B power supply. The 66111A also has an Agilent proprietary feature that helps test engineers avoid improper remote voltage settings and guarantees correct calibration of charger circuitry of the device under test.
| 66111A Specifications |
| Number of Outputs |
1 |
Output Ratings
(0-55°C) |
0-15 V, 0-3 A
Peak current for up to 7 ms: 5A |
Programming Accuracy
(25°C ± 5°C) |
0.05% + 10 mV
0.05% + 1.33 A |
| Ripple and Noise (in the range of 20 Hz to
20 MHz, with outputs ungrounded or with either terminal grounded) |
Voltage (rms/p-p) Normal Mode:
1 mV/6 mV
Current (rms): 2 mA |
| DC Measurement Accuracy (via GPIB or front
panel meters with respect to actual output at 25°C ± 5°C ) |
Voltage: 0.03% + 5 mV
(+20 mA to +rated I): 0.2% + 9 mA
(-20 mA to -rated I): 0.2% + 9 mA |
| Load Regulation |
Voltage: 2 mV
Current: 0.75 mA |
| Line Regulation |
Voltage: 0.5 mV
Current: 0.75 mA |
| Transient Response Time (for output to recover
to within 20 mV of final value after 0.1 to 1.5 A load change in high mode
compensation range) |
< 35 us |
| Supplemental Characteristics |
| Average Programming Resolution |
Voltage: 3.75 mV
Current: 0.67 mA |
| Sink current |
-2 A @ 7.5 V |
| Transient Voltage Dip (typical with up to
15 feet 22 AWG wiring) |
70 mV |
| DC Floating Voltage |
Output terminals can be floated
up to ±50 VDC maximum from chassis ground. |
| Remote Sensing |
Up to 4 volts can be dropped in
each load lead. Add 2 mV to the voltage load regulation specification for
each 1 volt change (in the positive output lead) due to load current change |
| Command Processing Time |
Average time required for the
output voltage to begin to change following receipt of digital data is 4
ms for power supplies connected directly to the GPIB. |
| Output Programming Response Time |
The rise and fall time (10/90%
and 90/10%) of the output voltage is less than 200 us. |
| Output Voltage Settling Time |
The output voltage change settles
within 1 LSB (0.025% X rated voltage) of final value in 2 ms. |
| Measurement Time |
Average time to make a voltage
or current measurement is 50 ms. |
| GPIB Interface Capabilities |
IEEE 488.2, SCPI command set and
Agilent 6630A Series programming compatibility. |
| Input Power (full load): 47-63 Hz |
115 VAC mains (104-127 Vac): 1.4
A, 100 W |
| Regulatory Compliance |
Listed to UL 3111-1, certified
to CSA C22.2 no. 1010.1-92; conforms to EN61010-1; conforms to IEC 1010-1,
complies with European EMC and Low Voltage directives, carries the CE mark.
Also complies with the Australian EMC requirements and carries the C-tick
mark. |
| Physical Specifications |
| Weight |
8.85 kg (19.5 lb) net
11.1 kg (24.5 lb) shipping |
| Size |
88.1 mm H x 212.8 mm W x 444.4
mm D (3.5 in x 8.4 in x 17.5 in ) |