TestEquity Model 3007C FastRate™ Temperature Chamber provides fast temperature transition rates in a compact, self-contained package. Dual 3.5 HP scroll compressors, boost heaters, and two heavy-duty blower motors give you the extra capacity you need for ramp rates in excess of 10°C/minute without liquid nitrogen (LN2). If you need even more cooling capacity, LN2 boost cooling capability is available.More Effective Reliability Screening Faster Test Time |
Programmable
Temperature Controller
Up to 256 steps can be programmed into as many as 40 namable profiles. The
context sensitive information key and guided steps make profile programming
fast and easy. A four-line backlit LCD displays programming, setup, operating
and help information. A large LED readout indicates the actual chamber temperature
with 0.1° resolution. Internal logic provides refrigeration compressor
control for responsive and reliable performance. Includes two alarms and seven
event outputs to control remote devices. RS-232C communications is also included.
GPIB, Ethernet and analog retransmit are available as options.
More about the Temperature Controller »
RS-232 Programming Resources »
GPIB Programming Resources and LabVIEW drivers »
Ethernet Interface Information »
High/Low Temperature Limit Controller and Alarm
Provides
independent protection against excess temperatures. Both high and low limits
can be set. The upper LED readout indicates the actual chamber temperature
while the lower readout indicates the status The limit controller will shut
down the chamber and trigger the audible alarm in the event of an out of limit
condition. A set of uncommitted contacts are available to provide a safety
power interlock for your test sample.
Heavy Duty Blowers
The system includes two forward curved stainless-steel, welded-construction blower
wheels in a counter rotating pattern to provide high velocity airflow throughout
the workspace, driven by two heavy-duty motors.
Viewing Window
Our unique design virtually eliminates exterior window sweating during normal
temperature cycling conditions.
Event Relays Standard
Three uncommitted solid-state relays are provided to switch the power to remote
AC devices via the F4 Temperature Controller.
Two 4" Access Ports
Lets you attach wires and sensors to your test sample through both the left
and right side of the chamber.
Adjustable Shelf
A heavy-duty stainless steel wire shelf ensures proper airflow around your
test sample for uniform temperature distribution.
Chamber Condensate Drain
The chamber floor is sloped and has a drain to permit removal of condensate
that can occur when cycling from low to high temperatures.
Casters
Rugged 3" casters permit easy movement.
Built-In Refrigeration Gauges
Provides system status for easy preventive maintenance and service.
World's
Best Environmental Chamber Warranty
Only TestEquity offers a 3-year warranty on parts and 1-year warranty
on labor. Our nationwide service network is qualified to do the job right.
And we always have replacement parts in stock for immediate shipment. Labor
warranty applies to the United States and Canada only. See the complete warranty
statement here »
LN2 (Liquid Nitrogen) Boost Capability (optional)
Provides extra cooling capacity for higher ramp rates and increased load capacity.
For example, cooling with compressors only from +85°C to -40°C in
an empty chamber is at a maximum rate of 12°C/minute. With optional LN2
Boost, the cooling rate for that condition increases to 20°C/minute.
Or, with a 25 pound load, the rate increases from 8°C/minute with compressors
only to 12°C/minute with LN2 Boost.
The system uses a specially-designed cryogenic solenoid valve to regulate the flow of LN2 in the chamber. A redundant cryogenic solenoid valve provides low-temperature protection in the event of a control valve failure, a controller failure, or operator error. The valves are located in a double-insulated sealed compartment to minimize thermal losses and prevent moisture condensation.
No-Sweat Cabinet
We use a combination of structural foam and fiberglass blanket insulation for
a higher combined R-factor than just fiberglass alone. The door sealing surface
is kept warm by circulating excess heat from the refrigeration system. The
result is a cabinet that does not sweat due to condensation during low temperature
operation.
The window is constructed of 6 panes of tempered glass separated by low-conductivity spacers to keep thermal losses to a minimum. Our unique design virtually eliminates exterior window sweating during normal temperature cycling conditions.
The stainless-steel interior liner is heliarc welded, while the refrigeration compartment behind the liner is double-insulated and carefully sealed. These design features reduce thermal losses and prevent moisture migration, ensuring years of optimum performance while keeping the cabinet rust-free.
Reliability & Safety That's Designed In
Non-CFC refrigerants are used in a cascade (two compressors) configuration.
Expansion valves provide optimum performance throughout the operating range.
Pressure regulators keep the operating pressures within safe limits regardless
of load conditions. Liquid injection ensures cool compressor operation during
high-temperature pull down for long life. Sequential starting of each compressor
reduces the current demand on start-up. Each compressor is protected by high/low
pressure switches. Care has been taken to use a minimum number of fittings,
joints and welds to reduce the possibility of refrigerant leaks.
All electrical control components are UL approved. Each branch circuit is individually protected with class-CC fuses. A main power disconnect switch is provided on the front panel. The nichrome air-heater has a low watt-density and even heat distribution for reliable performance and fast response. A fusible link provides fail-safe protection against thermal runaways, in addition to the microprocessor-based high/low limit controller.
The side and rear panels are removable to permit complete access to the refrigeration system for ease of service and preventive maintenance.
About Copeland Glacier Compliant Scroll® Compressors
Compliant Scroll is the most reliable, most efficient and quietest compressor
technology available today. Copeland's design allows the compressor to be
more tolerant of liquid refrigerant and debris. Compliant Scroll compressors
perform at higher efficiency levels than reciprocating compressors. They
also have 50% fewer moving parts, which increases reliability and reduces
sound levels. More about Scroll Compressors »
|
|
|
TestEquity 3007C FastRate Temperature Chamber (Environmental Chamber) Includes programmable controller with RS-232 interface, high/low limit controller & alarm, window, interior light, adjustable shelf, 4" port on left and right side, four AC event relays, chamber drain, built-in refrigeration gauges, casters, NIST traceable calibration certificate and manual. Input voltage must be specified at time of order. Specify 208V, 3 PH (TE-0724) or 230V, 3 PH (TE-0722) (3 Phase input only). |
TestEquity Price $27,995
|
|
|
|
||
|
0001 Circular Chart Recorder Honeywell DR4300 Series. One pen, 10" circular chart. |
TestEquity Price $995
|
||
|
|
|||
|
0003 GPIB Interface (Factory Installed Only) Converts the temperature controller's serial interface to GPIB. (order GPIB Option 1052 for short-term rentals) More info >> |
TestEquity Price $895
|
||
|
|
|||
|
0004 Controller Retransmit Outputs (Installed) Analog retransmit of chamber temperature, set point (or %RH on 123H and 1007H). Two outputs. More info >> |
TestEquity Price $195
|
||
|
|
|||
|
0019 Dry Air Purge (Factory Installed Only) Heatless regenerative desiccant air drier (installed). Requires 8 cfm supply air flow rate at 70°F and 100 psig (150 psig max) to be supplied by customer. More info >> Not available for rentals. |
TestEquity Price $2,995
|
||
|
|
|||
|
0009 GN2 Purge (Factory Installed Only) Requires GN2 to be supplied by customer. More info >> Not available for rentals. |
TestEquity Price $995
|
||
|
|
|||
|
0018 LN2 Boost Cooling Liquid Nitrogen (LN2) boost provides extra cooling capacity. Requires LN2 to be supplied by customer. |
TestEquity Price $1,995
|
||
|
|
|||
|
1052 GPIB Interface (External interface converter box) Converts the temperature controller's serial interface to GPIB. More info >> |
TestEquity Price $895
|
||
|
|
|||
|
1055 Ethernet Interface (External interface converter box) Converts the temperature controller's serial interface to Ethernet. More info >> |
TestEquity Price $895
|
||
|
|
|||
|
0040 Additional 4" Port (Factory Installed Only) Specify location at time of order. Chamber comes standard with one port on the left and right side (2 total). Additional ports not available for rentals. |
TestEquity Price $249
|
||
|
|
|||
|
0701 Extra Shelf with Clips for Model 1007 or 3007 Chamber comes standard with one shelf. |
TestEquity Price $95
|
||
|
|
|||
|
300534 Port Plug, Silicone Foam, 4-inch Open-cell silicone sponge with closed-cell silicone facing. |
TestEquity Price $48
|
||
|
|
|||
|
TestEquity 1007C Temperature Chamber (Environmental Chamber) 7 Cu Ft, -73°C to +175°C |
|
|
|
|
|
TestEquity Spare Parts For TestEquity Environmental Chambers |
|
|
|
|