Model 1027S Temperature Chamber Specifications
<< Back to Model 1027S Product Page
| Temperature Range |
-35°C to +175°C |
| Control Tolerance |
±0.5°C (Measured at the
control sensor after stabilization) |
| Uniformity |
±1.0°C (Variations throughout
the chamber after stabilization) |
| Cool Down Transition Time
(empty)* |
| Start
Temp |
End Temp |
| +23°C |
0°C |
-10°C |
-20°C |
-30°C |
-35°C |
| +150°C |
9.0 min |
14 min |
17 min |
20 min |
26 min |
Ultimate |
| +85°C |
4.0 min |
7.5 min |
10 min |
13 min |
19 min |
| +23°C |
----- |
1 min |
2 min |
4 min |
9 min |
| |
| Cool Down Transition Time (with 80
lb. aluminum load)* |
| Start
Temp |
End Temp |
| +23°C |
0°C |
-10°C |
-20°C |
-30°C |
-35°C |
| +150°C |
20.0 min |
28 min |
33 min |
38 min |
46 min |
Ultimate |
| +85°C |
8.0 min |
16 min |
20 min |
25 min |
33 min |
| |
| Heat Up Transition Time (empty)*
|
| Start
Temp |
End Temp |
| +23°C |
+50°C |
+85°C |
+125°C |
+150°C |
+175°C |
| +23°C |
----- |
1.5 min |
7 min |
14 min |
20 min |
25 min |
| 0°C |
1.5 min |
3.5 min |
13 min |
20 min |
23 min |
31 min |
| -10°C |
2.2 min |
4.2 min |
14 min |
22 min |
25 min |
33 min |
| -35°C |
5.5 min |
10 min |
16 min |
23 min |
29 min |
34 min |
| |
| Heat Up Transition Time (with 80 lb.
aluminum load)* |
| Start
Temp |
End Temp |
| +23°C |
+50°C |
+85°C |
+125°C |
+150°C |
+175°C |
| 0°C |
5 min |
13 min |
23 min |
36 min |
45 min |
55 min |
| -35°C |
11 min |
19 min |
29 min |
42 min |
51 min |
61 min |
|
Rate Of Change
To calculate rate of change for a particular condition, take the difference
between the Start Temp and End Temp and divide by the Transition Time.
Cool Down Example (empty): From +85°C to 0°C = 85°C
/ 7.5 min = 11.3°C/min.
Cool Down Example (with 80 lb. load): From +85°C to 0°C
= 85°C / 16 min = 5.3°C/min.
Heat Up Example (empty): From 0°C to +85°C = 85°C
/ 13 min = 6.5°C/min.
Heat Up Example (with 80 lb. load): From 0°C to +85°C
= 85°C / 23 min = 3.7°C/min.
*Note: Transition times are measured after a 2 hour soak at
the respective start temperature at 208 V input. |
| Live Load
Capacity |
| +23°C |
0°C |
-10°C |
-20°C |
-30°C |
| 4,000 Watts |
2,800 Watts |
2,300 Watts |
1,800 Watts |
1,250 Watts |
| Refrigeration
and Heating System |
| Refrigerant |
R-404A (Dupont HP-62) |
| Compressors |
3.5 HP Copeland scroll compressor. More
about Scroll Compressors >> |
| Condenser |
Air Cooled |
| Heat of Rejection |
30,200 BTUH (maximum rated chamber load at maximum
cooling rate from high temperature soak) |
| Heater Power |
4,000 Watts @ 208 V input |
| |
| Instrumentation |
| Temperature Controller |
256 step, 40 profile, ramp and soak programmable memory.
RS-232/485 interface. More details >>
|
 |
| Limit Controller |
Independent high and low temperature limits. Triggers an
audible alarm and shuts down the chamber. Relay contacts provide a safety
power interlock for test sample. |
| Chart Recorder |
(Optional) Honeywell DR4300 Series. One pen, 10" circular
chart. Mounts in lower front door. |
 |
| Input Power
Requirements |
| 230 V ±10%, 60 Hz, 3 Phase |
Max Current Draw 31 A; Recommended Service
40 A |
| 208 V -5/+10%, 60 Hz, 3 Phase |
Max Current Draw 26 A; Recommended Service
35 A |
| |
Input may be configured for 230 V or 208
V in the field by changing jumpers. Three phase load is balanced. Call
for other voltages or 50 Hz operation.
Customer power source must be hard-wired to the chamber by a qualified
electrician. Power cord and plug is not included. |
| |
| Physical
Characteristics and Safety |
| Inside Dimensions |
40" W x 32" H x 36.5" D (27 cubic
feet)
1016 mm W x 812.8 mm H x 927 mm D (764 liters) |
| Outside Dimensions |
49" W x 73.25" H x 63" D (nominal)
1244 mm W x 1860 mm H x 1600 mm D
Door latch adds 3" to width on right side. Circulator motor and housing
adds 6" to height - may be removed for move-in. |
 |
| Minimum Installed Clearance |
18" from the left and right side
24" from the rear |
| Window Viewing Area |
16" W x 13" H |
| Access Ports |
4" Port on left and right side (two total)
Supplied with foam plugs |
| Weight |
Chamber Weight: 1,600 pounds
Shipping Weight: 2,000 pounds |
| NOTE: Performance is typical and based on operation at 23°C
(73°F) ambient and nominal input voltage. Designed for use in a normal
conditioned laboratory. Operation at higher ambient temperatures may result
in decreased cooling performance. Additional ports and shelves will also
affect performance. Operation above 30°C (85°F) or below 16°C
(60°F) ambient is not recommended.
Due to continuous product improvements, specifications subject to change without notice.
|