Model 1007H Temperature/Humidity Chamber Specifications
<< Back to Model 1007H Product Page
| Temperature Range |
-73°C to +175°C |
| Control Tolerance |
±0.2°C (Measured at the control sensor
after stabilization) |
| Uniformity |
±0.5°C (Variations throughout the chamber
after stabilization) |
| |
| Humidity Range |
Standard Range: 10% to 95%
(Limited by a 6°C dewpoint and maximum dry bulb of +85°C)
With optional GN2 Purge or Dry Air Purge: 5% to 95%
(Dry bulb range of +10°C to +85°C)

|
| Control Tolerance |
±3% RH |
| Display Resolution |
0.1% RH |
| Humidity Sensor |
Dynamic capacitive type (no wet wicks required) |
| Cool Down Transition Time* (uncontrolled humidity
mode) |
|
Start Temp |
End Temp |
| +23°C |
0°C |
-40°C |
-55°C |
-65°C |
-73°C |
| +23°C |
----- |
4 min |
18 min |
25 min |
33 min |
Ultimate |
| +50°C |
5 min |
11 min |
27 min |
36 min |
45 min |
Ultimate |
| +85°C |
13 min |
20 min |
37 min |
47 min |
55 min |
Ultimate |
| +150°C |
32 min |
39 min |
58 min |
65 min |
76 min |
Ultimate |
| |
| Heat Up Transition Time* (uncontrolled humidity mode) |
|
Start Temp |
End Temp |
| +23°C |
+50°C |
+85°C |
+125°C |
+150°C |
+175°C |
| +23°C |
----- |
1.5 min |
7 min |
14 min |
20 min |
25 min |
| 0°C |
1.5 min |
3.5 min |
13 min |
20 min |
23 min |
31 min |
| -40°C |
6 min |
11 min |
17 min |
24 min |
30 min |
35 min |
| -55°C |
8 min |
13 min |
19 min |
26 min |
32 min |
37 min |
| -65°C |
10 min |
14 min |
21 min |
28 min |
34 min |
39 min |
| Rate Of Change
To calculate rate of change for a particular condition, take the difference
between the Start Temp and End Temp and divide by the Transition Time.
Cool Down Example: From +50°C to -40°C = 90°C / 27
min = 3.33°C/min.
Heat Up Example: From -40°C to +50°C = 90°C / 11 min
= 8.18°C/min.
*Note: Transition times are measured after a 2 hour soak at the
respective start temperature with an empty chamber. |
| Live Load
Capacity (uncontrolled humidity mode) |
| +23°C |
0°C |
-40°C |
-55°C |
-65°C |
| 1,000 Watts |
800 Watts |
500 Watts |
400 Watts |
300 Watts |
| Refrigeration
and Heating System |
| High Stage Refrigerant |
R-404A (Dupont HP-62) |
| Low Stage Refrigerant |
R-508B (Dupont SUVA-95) |
| Compressors |
1 HP x 1 HP Copeland hermetic compressors in
a cascade configuration. |
| Condenser |
Air Cooled |
| Heat of Rejection |
14,800 BTUH (maximum rated chamber load at maximum
cooling rate from high temperature soak) |
| Air Heater Power |
2,000 Watts |
| Humidifier Heater Power |
1,500 Watts (rating at 240V) |
| |
| Instrumentation |
| Temp/Humidity Controller |
256 step, 40 profile, ramp and soak programmable memory.
RS-232/485 interface. More details >>
|
 |
| Limit Controller |
Independent high and low temperature limits. Triggers an
audible alarm and shuts down the chamber. Relay contacts provide a safety
power interlock for test sample. |
| Chart Recorder |
(Optional) Honeywell DR4300 Series. Two pen, 10" circular
chart. Mounts in lower front door. 24-hour rotation. |
 |
| Input Power Requirements |
| |
Phase |
Current Draw |
Minimum Service |
| 230 V ±10%, 60 Hz |
1 PH |
25 A |
30 A |
| 3 PH |
19 A |
25 A |
| |
| 208 V -5/+10%, 60 Hz |
1 PH |
28 A |
35 A |
| 3 PH |
21 A |
30 A |
| Input may be configured for single or three
phase in the field by changing jumpers. Three phase load is semi-balanced.
Call for other voltages or 50 Hz operation. Cooling performance is reduced
17% at 50 Hz.
Customer power source must be hard-wired to the chamber by a qualified
electrician. Power cord and plug is not included. |
| Humidity
Water Requirements |
| Supply and Drain |
Must be provided with a water line and floor drain. Negligible
consumption. |
| Water Recirculation System (optional) |
Provides a reliable supply of filtered water for the humidity
system. Perfect for installations where a water line and drain are not available.
|
| Physical
Characteristics and Safety |
| Inside Dimensions |
24" W x 21" H x 24" D (7 cubic feet)
609 mm W x 533 mm H x 609 mm D (198 liters) |
| Outside Dimensions |
33" W x 68.75" H x 54" D (nominal)
838 mm W x 1746 mm H x 1372 mm D |
| Door latch adds 3" to width on right side
(may be removed to permit move-in through a 36" doorway). Circulator
motor and housing adds 6" to height. |
| Minimum Installed Clearance |
18" from the left and right side
24" from the rear |
| Window Viewing Area |
13" W x 13" H |
| Access Ports |
4" Port on left and right side (two total)
Supplied with foam plugs |
| Weight |
Chamber Weight: 900 pounds
Shipping Weight: 1,050 pounds |
| Sound Level |
62 dBA in cooling mode (A-weighted, measured 36" from
the front or side surface, 63" from the floor, in a free-standing environment) |
| NOTE: Performance is typical and based on operation at 23°C
(73°F) ambient and nominal input voltage. Designed for use in a normal
conditioned laboratory. Operation at higher ambient temperatures may result
in decreased cooling performance. Additional ports and shelves will also
affect performance. Operation above 30°C (85°F) or below 16°C
(60°F) ambient is not recommended. |
<< Back to Model 1007H Product Page
|