TestEquity Inc - http://www.testequity.com/ To order, call us at 800-950-3457
or visit us on the web: www.TestEquity.com

Fluke 724


Summary Specifications (18°C to 28°C for one year) *
Voltage
(max load, 1 mA)

0 to 100 mV:
Resolution: 0.01 mV
Accuracy: 0.02% Rdg + 2 LSD

0 to 10V (source):
Resolution: 0.001V
Accuracy: 0.02% Rdg + 2 LSD

0 to 30V (measure):
Resolution: 0.001V
Accuracy: 0.02% Rdg + 2 LSD

mA
(max load, 1000 Ω)

0 to 24 mA:
Resolution:0.001 mA
Accuracy: 0.02% Rdg + 2 LSD

mV
(TC terminals)

-10.00 mV to +75.00 mV:
Resolution: 0.01 mV
Accuracy: .025% Rdg + 1 LSD

Resistance

0 Ω to 3200 Ω (Measure):
Resolution: 0.01 Ω to 0.1 Ω
Accuracy: 0.10 Ω to 1.0 Ω

15 Ω to 3200 Ω (Source):
Resolution: 0.01 Ω to 0.1 Ω
Accuracy: 0.10 Ω to 1.0 Ω

Loop Supply 24V dc:
Resolution: N/A
Accuracy: 10%
*Temperature coefficient, -10°C to 18°C, 28°C to 55°C, ±0.005% of range per °C.
 
Thermocouple Accuracy Specifications (measure or source) *
J

-200 to 0°C:
Accuracy: 1.0°C

0 to 1200°C:
Accuracy: 0.7°C

K

-200 to 0°C:
Accuracy: 1.2°C

0 to 1370°C:
Accuracy: 0.8°C

T

-200 to 0°C:
Accuracy: 1.0°C

0 to 400°C:
Accuracy: 0.8°C

E

-200 to 0°C:
Accuracy: 0.9°C

0 to 950°C:
Accuracy: 0.7°C

R

-20 to 0°C:
Accuracy: 2.5°C

0 to 500°C:
Accuracy: 1.8°C

500 to 1750°C:
Accuracy: 1.4°C

S

-20 to 0°C:
Accuracy: 2.5°C

0 to 500°C:
Accuracy: 1.8°C

500 to 1750°C:
Accuracy: 1.5°C
B

600 to 800°C:
Accuracy: 2.2°C

800 to 1000°C:
Accuracy: 1.8°C

1000 to 1800°C:
Accuracy: 1.4°C

L

-200 to 0°C:
Accuracy: 0.85°C

0 to 900°C:
Accuracy: 0.7°C

U

-200 to 0°C:
Accuracy: 1.1°C

0 to 400°C:
Accuracy: 0.75°C

N

-200 to 0°C:
Accuracy: 1.5 °C

0 to 1300°C:
Accuracy: 0.9°C

* (Accuracy specifications include 0.2°C cold junction uncertainty)

Thermocouple Resolution
J, K, T, E, L, N, U 0.1°C, 0.1°F
B, R, S 1°C, 1°F
 
RTD Ranges and Accuracy Specifications *
Ni 120

-80°C to 260°C:
Measure: 0.2°C
Source: 0.2°C

Pt 100 - 385 -200°C to 800°C:
Measure: 0.33°C
Source: 0.33°C
Pt 100 - 3926 -200°C to 630°C:
Measure: 0.3°C
Source: 0.3°C
Pt 100 - 3916 (JIS) -200°C to 630°C:
Measure: 0.3°C
Source: 0.3°C

Pt 200 - 385

-200°C to 250°C:
Measure: 0.2°C
Source: 0.2°C

250°C to 630°C:
Measure: 0.8°C
Accuracy: 0.8°C

Pt 500 - 385

-200°C to 500°C:
Measure: 0.3°C
Source: 0.3°C

500°C to 630°C:
Measure: 0.4°C
Accuracy: 0.4°C

Pt 1000 - 385

-200°C to 100°C
Measure: 0.2°C
Source: 0.2°C

100°C to 630°C
Measure: 0.3°C
Source: 0.2°C

* Note: Measure (4 wire)

RTD Resolution 0.1°C, 0.1°F
 
General Specifications
Maximum voltage 30V
Storage Temperature -20°C to 71°C
Operating Temperature -10°C to 55°C
Relative Humidity 90% (10°C to 30°C); 75% (30°C to 40°C);
45% (40°C to 50°C); 35% (50°C to 55°C)
Vibration Random, 2g, 5-500 Hz
Safety CSA C22.2 No. 1010.1:1992
EMC EN50082-1:1992 and EN55022:1994 Class B
Size 96 x 200 x 47 mm (3.8 x 7.9 x 1.9 inches)
Weight 650g (23oz)
Battery Four AA alkaline batteries. Battery life: 25 hours typical
Warranty Three years

 

 
TestEquity Inc - http://www.testequity.com/ To order, call us at 800-950-3457