Model 1027C Temperature Chamber Specifications
<< Back to Model 1027C Product Page
| Temperature Range |
-73°C to +175°C |
| Control Tolerance |
±0.2°C (Measured at the control
sensor after stabilization) |
| Uniformity |
±1.0°C (Variations throughout
the chamber after stabilization) |
| Cool Down Transition Time (empty)* |
| Start
Temp |
End Temp |
| +23°C |
0°C |
-20°C |
-40°C |
-55°C |
-65°C |
| +23°C |
----- |
1.5 min |
7.5 min |
11 min |
14 min |
18 min |
| +85°C |
9 min |
15 min |
21 min |
28 min |
35 min |
40 min |
| +150°C |
23 min |
30 min |
37 min |
45 min |
53 min |
60 min |
| |
| Cool Down Transition Time (with 80 lb.
aluminum load)* |
| Start
Temp |
End Temp |
| +23°C |
0°C |
-20°C |
-40°C |
-55°C |
-65°C |
| +85°C |
15 min |
26 min |
35 min |
48 min |
59 min |
68 min |
| +150°C |
32 min |
41 min |
50 min |
61 min |
69 min |
76 min |
| |
| Heat Up Transition Time (empty)* |
| Start
Temp |
End Temp |
| +23°C |
+50°C |
+85°C |
+125°C |
+150°C |
+175°C |
| +23°C |
----- |
1.5 min |
7 min |
14 min |
20 min |
25 min |
| 0°C |
1.5 min |
3.5 min |
13 min |
20 min |
23 min |
31 min |
| -40°C |
6 min |
11 min |
17 min |
24 min |
30 min |
35 min |
| -55°C |
8 min |
13 min |
19 min |
26 min |
32 min |
37 min |
| -65°C |
10 min |
14 min |
21 min |
28 min |
34 min |
39 min |
| |
| Heat Up Transition Time (with 80 lb.
aluminum load)* |
| Start
Temp |
End Temp |
| +23°C |
+50°C |
+85°C |
+125°C |
+150°C |
+175°C |
| 0°C |
5 min |
13 min |
23 min |
36 min |
45 min |
55 min |
| -40°C |
11 min |
19 min |
29 min |
42 min |
51 min |
61 min |
| -65°C |
22 min |
32 min |
44 min |
61 min |
71 min |
81 min |
Rate Of Change
To calculate rate of change for a particular condition, take the difference
between the Start Temp and End Temp and divide by the Transition Time.
Cool Down Example (empty): From +85°C to -40°C = 125°C
/ 28 min = 4.5 °C/min.
Cool Down Example (with 80 lb. load): From +85°C to -40°C
= 125°C / 48 min = 2.6°C/min.
Heat Up Example (empty): From -40°C to +85°C = 125°C
/ 17 min = 7.35°C/min.
Heat Up Example (with 80 lb. load): From -40°C to +85°C
= 125°C / 29 min = 4.3°C/min.
*Note: Transition times are measured after a 2 hour soak at
the respective start temperature at 208 V input. |
| Live
Load Capacity |
| +23°C |
0°C |
-20°C |
-40°C |
-55°C |
-65°C |
| 2,900 Watts |
2,600 Watts |
2,300 Watts |
1,750 Watts |
1,450 Watts |
1,050 Watts |
| Refrigeration and Heating
System |
| High Stage Refrigerant |
R-404A (Dupont HP-62) |
| Low Stage Refrigerant |
R-508B (Dupont SUVA-95) |
| Compressors |
3.5 HP x 3.5 HP Copeland scroll compressors
in a cascade configuration
More about Scroll Compressors >> |
| Condenser |
Air Cooled |
| Heat of Rejection |
27,500 BTUH (maximum rated chamber load at maximum
cooling rate from high temperature soak) |
| Heater Power |
4,000 Watts @ 208 V input |
| |
| Instrumentation |
| Temperature Controller |
256 step, 40 profile, ramp and soak programmable memory.
RS-232/485 interface.
More details >> |
 |
| Limit Controller |
Independent high and low temperature limits. Triggers an
audible alarm and shuts down the chamber. Relay contacts provide a safety
power interlock for test sample. |
| Chart Recorder |
(Optional) Honeywell DR4300 Series. One pen, 10" circular
chart. Mounts in lower front door. |
 |
| Input Power
Requirements |
| 230 V ±10%, 60 Hz, 3 Phase |
Max Current Draw 39 A; Recommended Service 50
A |
| 208 V -5/+10%, 60 Hz, 3 Phase |
Max Current Draw 35 A; Recommended Service 45
A |
| |
Input may be configured for 230 V or 208
V in the field by changing jumpers. Three phase load is balanced. Call
for other voltages or 50 Hz operation.
Customer power source must be hard-wired to the chamber by a qualified
electrician. Power cord and plug is not included. |
| |
| Physical
Characteristics and Safety |
| Inside Dimensions |
40" W x 32" H x 36.5" D (27 cubic
feet)
1016 mm W x 813 mm H x 927 mm D (765 liters) |
| Outside Dimensions |
49" W x 73.25" H x 63" D (nominal)
1244 mm W x 1860 mm H x 1600 mm D
Door latch adds 3" to width on right side. Circulator motor and housing
adds 6" to height - may be removed for move-in. |
 |
| Minimum Installed Clearance |
18" from the left and right side
24" from the rear |
| Window Viewing Area |
16" W x 13" H |
| Access Ports |
4" Port on left and right side (two total)
Supplied with foam plugs |
| Weight |
Chamber Weight: 1,600 pounds
Shipping Weight: 2,000 pounds |
| NOTE: Performance is typical and based on operation at 23°C
(73°F) ambient and nominal input voltage. Designed for use in a normal
conditioned laboratory. Operation at higher ambient temperatures may result
in decreased cooling performance. Additional ports and shelves will also
affect performance. Operation above 30°C (85°F) or below 16°C
(60°F) ambient is not recommended.
Due to continuous product improvements, specifications subject to change
without notice. |