Cool Down Transition Time (empty)* | |||||||||
Start Temp | End Temp | ||||||||
+23°C | 0°C | -10°C | -20°C | -30°C | -35°C | -40°C | -42°C | ||
+23°C | Standard | ---- | 7 min | 11 min | 15 min | 19 min | 23 min | 30 min | Ultimate |
50 Hz Export Version | ---- | 8 min | 13 min | 18 min | 23 min | 28 min | 36 min | ||
+85°C | Standard | 13 min | 20 min | 24 min | 28 min | 32 min | 36 min | 45 min | Ultimate |
50 Hz Export Version | 16 min | 24 min | 29 min | 33 min | 38 min | 43 min | 54 min | ||
Rate Of Change *Note: Transition times are measured after a 30 minute soak at the respective start temperature with an empty chamber, as indicated on the temperature controller, 23°C (73°F) ambient. Measured with setpoint beyond the start and end temperatures. Does not include the effect of proportional band when approaching setpoint. |
Live Load Capacity | ||||
+23°C | 0°C | -30°C | -40°C | |
200 Watts | 155 Watts | 100 Watts | 35 Watts |
Refrigeration and Heating System | ||
Compressor | 1/3 HP Copeland hermetic | |
Condenser | Air Cooled | |
Heat of Rejection | 3,000 BTUH (maximum rated chamber load at maximum cooling rate from high temperature soak) | |
Heater Power | 500 Watts | |
Instrumentation | ||
Temperature Controller | Watlow F4T Touch Screen Controller with RS-232, Ethernet interface, 4.3" color graphic touch screen. OR... Watlow F4 Controller with RS-232 interface, LED readout of temperature, LCD display of other parameters. |
|
Limit Controller | Independent of temperature controller. User adjustable high and low temperature limits. Shuts down the chamber if limits are exceeded. Watlow EZ-Zone Series. |
Power Requirements | |
Input Voltage | Standard Model 106 120 V nominal (110 to 126 VAC), 60 Hz, 1 PH Max Current Draw 10 A, Recommended Minimum Service 15 A |
Export Model 106-EX 230 V nominal (209 to 253 VAC), 50 Hz, 1 PH Max Current Draw 5 A, Recommended Minimum Service 10 A |
|
Physical Characteristics | |
Inside Dimensions | 9" W x 9" H x 13.25" D, 0.62 cubic feet (229 x 229 x 337 mm, 17 liters) |
Outside Dimensions (nominal) |
23" W x 26" H x 16" D (584 x 660 x 406 mm) Depth with door 16.2" (498 mm) |
Minimum Installed Clearance | 12" (304 mm) from the rear |
Access Ports | 4 " (3.83" inside diameter) Port on left side Supplied with silicone foam plug |
Weight | Chamber Weight: 124 pounds (56 kg) Shipping Weight: 158 pounds (71 kg) |
Sound Level | 52 dBA in cooling mode (A-weighted, measured 36" from the front surface, 63" from the floor, in a free-standing environment) |
NOTE: Performance is typical and based on operation at 23°C (73°F) ambient and nominal input voltage. Designed for use in a normal conditioned laboratory. Operation at higher ambient temperatures will result in decreased cooling performance. Low end limit derates to -38°C when operating above 27°C (80°F) ambient. Operation above 30°C (85°F) or below 16°C (60°F) ambient is not recommended.