TestEquity Model TEC1 Thermoelectric Chamber Specifications


Temperature Range
+10°C to +60°C
Control Tolerance
±0.1°C (Measured at the control sensor after stabilization)
Uniformity
±1.0°C (Variations throughout the chamber after stabilization, at least 2-inches from any inside surface, in the range of +15°C to +55°C)
Accuracy
±2.0°C (Indication on temperature controller compared to measurement in the center of the chamber, in the range of +15°C to +55°C, after 30-minute stabilization at set point)
Temperature Sensor
Platinum RTD sensor, Class A accuracy
Temperature Controller
Watlow EZ-Zone, PID control, autotuning, ramp/soak programming with up to four profiles and 40 total steps, 0.1°C resolution, RS-232 interface. (optional GPIB & Ethernet converters)
 

Cool Down Time


Start Temperature End Temperature
+37°C +23°C +15°C +10°C
+55°C 7 min 16 min 26 min Ultimate
+23°C -- -- 7 min Ultimate
Heat Up Time
Start Temperature End Temperature
+23°C +37°C +55°C +60°C
+23°C -- 3.5 min 10 min Ultimate
+15°C 2.0 min 5 min 11 min Ultimate

*Note: Heat and cool transition times are measured in an empty chamber after a 30 minute soak at the start temperature, as indicated on the temperature controller. Transition times do not include the effect of proportional band when approaching the temperature set point.


Live Load Capacity
+37°C
+23°C
+15°C
+10°C
90 Watts
50 Watts
35 Watts
20 Watts
Workspace Dimensions
16" W x 8" H x 13" D (0.96 cubic feet)
Outside Dimensions
24.7" W x 12" H x 20.4" D (nominal)
Min. Installed Clearance
4" from the rear
Cable Slot
1" x 3" Cable Slot on left and right side (two total)
Weight
52 pounds. Shipping weight: 68 pounds.
Sound Level
Less than 50 dBA (A-weighted, measured 36" from the front surface, in a free-standing environment)
Input Power
88 to 132VAC / 176 to 264VAC, 50/60 Hz (internal switch for voltage range)
6A at 115 VAC, 3.5A at 230 VAC

NOTE: Performance is typical and based on operation at 23°C (73°F) ambient and nominal input voltage. This product is designed for use in a normal conditioned laboratory. Operation at higher ambient temperatures will result in decreased cooling performance. Operation above 27°C (80°F) will have a significant impact on low-temperature operation.


Due to continuous product improvement, specifications are subject to change without notice.