Available For Rent
- Description
- Specifications
- Documents
Keysight DP0030A High Voltage Differential Probe
Features and Specifications
- Bandwidth: 100 MHz
- Attenuation (high/low): 250/25:1
- Max differential measurement range (DC + AC peak): <750 / <75 V
- Oscilloscope interface: BNC + USB
- Differential input resistance: 5 MΩ / 2 pF
- Output cable length: 1.2 m
- Safety Rating(s): 300 Vrms CAT III / 600 Vrms CAT II
Keysight's DP0030A high-voltage differential active probe excels in general-purpose differential signal measurements, essential for high-speed power assessments like those of switching power devices, DC-DC converters, class D amplifiers, automotive bus measurements, and high-speed digital system designs. As designs become more intricate and the demand for floating measurements increases, this probe facilitates floating measurements between any two points on a board and can also operate as single-ended probes with one terminal grounded.
In addition, the DP0030A probe features a BNC interface and uses a USB for power delivery to the probe. It has a physical button to achieve switchable attenuation.
DP003XA Series of Probes
The DP003xA series of probes are compatible with most of Keysight’s InfiniiVision and Infiniium oscilloscopes that provide 1 megohm input terminations with the AutoProbe1 interface (DP0030A contains a BNC).
Series Models
- DP0030A - 100 MHz, <750 / <75 V (DC + AC peak)
- DP0031A - 200 MHz, <750 / <75 V (DC + AC peak)
- DP0032A - 200 MHz, <1500 / <150V (DC + AC peak)
- DP0033A - 200 MHz, <1500 / <150V (DC + AC peak)
- DP0034A - 70 MHz, <7000 / <700V (DC + AC peak)