Half Cube Model 105A Benchtop Temperature Chamber Specifications

Temperature Range
-40°C to +130°C
Control Tolerance
±0.5°C (±0.2°C Typical) (Measured at the control sensor after stabilization)
Uniformity
±1.0°C (Variations throughout the chamber after stabilization)

Cool Down Transition Time (empty)*
Start TempEnd Temp
+23°C0°C-10°C-20°C-30°C-35°C-40°C
+23°CStandard----3 min5 min7 min10 min13 min22 min
50 Hz Export Version----3 min5 min7 min11 min15 min24 min
+85°CStandard8 min13 min15 min17 min20 min23 min31 min
50 Hz Export Version9 min14 min16 min18 min21 min27 min41 min
Heat Up Transition Time (empty)*
Start TempEnd Temp
+23°C+50°C+85°C
+23°C-----1.5 min6 min
0°C2 min4 min7.5 min
-20°C3 min5 min8 min
-40°C4.5 min7.5 min11 min

Rate Of Change
To calculate rate of change for a particular condition, take the difference between the Start Temp and End Temp and divide by the Transition Time.
Cool Down Example (empty): From +85°C to -20°C = 105°C / 17 min = 6.18°C/min.
Heat Up Example: From -40°C to +85°C = 125°C / 11 min = 11.36°C/min.

*Note: Transition times are measured after a 30 minute soak at the respective start temperature.


Live Load Capacity
+23°C0°C-10°C-20°C-30°C-40°C
200 Watts175 Watts165 Watts145 Watts90 Watts10 Watts

Refrigeration and Heating System
Compressor1/3 HP Copeland hermetic
CondenserAir Cooled
Heat of Rejection3,000 BTUH (maximum rated chamber load at maximum cooling rate from high temperature soak)
Heater Power500 Watts
Instrumentation
Temperature Controller256 steps, 10 profiles, ramp and soak programmable memory. RS-232C interface. Watlow Series F4.
More details >>
Limit ControllerIndependent of temperature controller. User adjustable high and low temperature limits. Shuts down the chamber if limits are exceeded.

Power Requirements
Input VoltageStandard Model 105A
120 V nominal (110 to 126 VAC), 60 Hz, 1 PH
Max Current Draw 10 A, Recommended Minimum Service 15 A
Export Model 105A-EX
230 V nominal (209 to 253 VAC), 50 Hz, 1 PH
Max Current Draw 5 A, Recommended Minimum Service 10 A
Physical Characteristics and Safety
Inside Dimensions12" W x 9" H x 8" D, 0.5 cubic feet
(305 x 229 x 203 mm, 14 liters)
Outside Dimensions
(nominal)
16.5" W x 26" H x 20" D
(419 x 660 x 508 mm)
Door latch adds 2" (51 mm) to width. Circulator motor housing adds 2" (51 mm) to depth in rear
Minimum Installed Clearance12" (304 mm) from the rear
Access Ports3" (2.83" inside diameter) Port on left and right side (two total)
Supplied with silicone foam plugs
WeightChamber Weight: 114 pounds (52 kg)
Shipping Weight: 140 pounds (64 kg)

NOTE: Performance is typical and based on operation at 23°C (73°F) ambient and nominal input voltage. Designed for use in a normal conditioned laboratory. Operation at higher ambient temperatures will result in decreased cooling performance. Low end limit derates to -38°C when operating above 27°C (80°F) ambient. Operation above 30°C (85°F) or below 16°C (60°F) ambient is not recommended.