| Cool Down Transition Time (empty)* | |||||||||
| Start Temp | End Temp | ||||||||
| +23°C | 0°C | -10°C | -20°C | -30°C | -35°C | -40°C | -42°C | ||
| +23°C | Standard | ---- | 7 min | 11 min | 15 min | 19 min | 23 min | 30 min | Ultimate |
| 50 Hz Export Version | ---- | 8 min | 13 min | 18 min | 23 min | 28 min | 36 min | ||
| +85°C | Standard | 13 min | 20 min | 24 min | 28 min | 32 min | 36 min | 45 min | Ultimate |
| 50 Hz Export Version | 16 min | 24 min | 29 min | 33 min | 38 min | 43 min | 54 min | ||
|
Rate Of Change *Note: Transition times are measured after a 30 minute soak at the respective start temperature with an empty chamber, as indicated on the temperature controller, 23°C (73°F) ambient. Measured with setpoint beyond the start and end temperatures. Does not include the effect of proportional band when approaching setpoint. | |||||||||
| Live Load Capacity | ||||
| +23°C | 0°C | -30°C | -40°C | |
| 200 Watts | 155 Watts | 100 Watts | 35 Watts | |
| Refrigeration and Heating System | ||
| Compressor | 1/3 HP Copeland hermetic | |
| Condenser | Air Cooled | |
| Heat of Rejection | 3,000 BTUH (maximum rated chamber load at maximum cooling rate from high temperature soak) | |
| Heater Power | 500 Watts | |
| Instrumentation | ||
| Temperature Controller | Watlow F4T Touch Screen Controller with RS-232, Ethernet interface, 4.3" color graphic touch screen. OR... Watlow F4 Controller with RS-232 interface, LED readout of temperature, LCD display of other parameters. | |
| Limit Controller | Independent of temperature controller. User adjustable high and low temperature limits. Shuts down the chamber if limits are exceeded. Watlow EZ-Zone Series. | |
| Power Requirements | |
| Input Voltage | Standard Model 106 120 V nominal (110 to 126 VAC), 60 Hz, 1 PH Max Current Draw 10 A, Recommended Minimum Service 15 A |
| Export Model 106-EX 230 V nominal (209 to 253 VAC), 50 Hz, 1 PH Max Current Draw 5 A, Recommended Minimum Service 10 A | |
| Physical Characteristics | |
| Inside Dimensions | 9" W x 9" H x 13.25" D, 0.62 cubic feet (229 x 229 x 337 mm, 17 liters) |
| Outside Dimensions (nominal) | 23" W x 26" H x 16" D (584 x 660 x 406 mm) Depth with door 16.2" (498 mm) |
| Minimum Installed Clearance | 12" (304 mm) from the rear |
| Access Ports | 4 " (3.83" inside diameter) Port on left side Supplied with silicone foam plug |
| Weight | Chamber Weight: 124 pounds (56 kg) Shipping Weight: 158 pounds (71 kg) |
| Sound Level | 52 dBA in cooling mode (A-weighted, measured 36" from the front surface, 63" from the floor, in a free-standing environment) |
NOTE: Performance is typical and based on operation at 23°C (73°F) ambient and nominal input voltage. Designed for use in a normal conditioned laboratory. Operation at higher ambient temperatures will result in decreased cooling performance. Low end limit derates to -38°C when operating above 27°C (80°F) ambient. Operation above 30°C (85°F) or below 16°C (60°F) ambient is not recommended.
Due to continuous product improvement, specifications are subject to change without notice.