Model 115 Temperature Chamber Specifications
Cool Down Transition Time* | ||||||
Start Temp | End Temp | |||||
+23°C | 0°C | -40°C | -55°C | -65°C | -73°C | |
+23°C | ----- | 5 min | 20 min | 30 min | 40 min | Ultimate |
+85°C | 16 min | 25 min | 45 min | 57 min | 69 min | Ultimate |
Rate Of Change |
||||||
*Note: Transition times are measured after a 1 hour soak at the respective start temperature with an empty chamber, as indicated on the temperature controller, 23°C ambient. |
Live Load Capacity | ||||
+23°C | 0°C | -40°C | -55°C | -65°C |
300 Watts | 255 Watts | 160 Watts | 110 Watts | 80 Watts |
Refrigeration and Heating System | |
High Stage Refrigerant | R-404A (Dupont HP-62) |
Low Stage Refrigerant | R-508B (Dupont SUVA-95) |
Compressors | 1/2 HP x 1/3 HP Copeland hermetic compressors in a cascade configuration. |
Condenser | Air Cooled |
Heat of Rejection | 4,000 BTUH (maximum rated chamber load at maximum cooling rate from high temperature soak) |
Heater Power | 600 Watts |
Instrumentation | ||
Temperature Controller | 256 step, 40 profile, ramp and soak programmable memory. RS-232/485 interface. More details >> |
![]() |
Limit Controller | Independent high and low temperature limits. Triggers an audible alarm and shuts down the chamber. |
Power Requirements | |
Input Voltage |
120 V nominal (110 to 126 VAC), 60 Hz, 1 PH Model 115-EX export version available with 230V, 50Hz, 8A input |
Power Cord and Plug |
Plugs into a standard NEMA 5-20R receptacle. Model 115-EX export version is supplied with power cord for destination country. |
Physical Characteristics and Safety | |
Inside Dimensions | 16" W x 12" H x 14" D (1.55 Cubic feet) 406 mm W x 305 mm H x 353 mm D (44 liters) |
Outside Dimensions |
Benchtop Model 115B Floor Model 115F |
Minimum Installed Clearance | 6" from the left and right side 12" from the rear |
Window Viewing Area | 8" W x 6" H |
Access Ports | 3" (2.83" inside diameter) supplied on all new Model 115 chambers. Port on left and right side (two total) Supplied with silicone foam port plugs |
Weight |
Benchtop Model 115B Floor Model 115F |
Sound Level | 52 dBA in cooling mode (A-weighted, measured 36" from the front surface, 63" from the floor, in a free-standing environment) |
CE Mark | Export version, Model 115-EX is CE Marked |
NOTE: Performance is typical and based on operation at 23°C (73°F) ambient and nominal input voltage. Designed for use in a normal conditioned laboratory. Operation at higher ambient temperatures may result in decreased cooling performance. Additional ports and shelves will also affect performance. Operation above 30°C (85°F) or below 16°C (60°F) ambient is not recommended.
Due to continuous product development, specifications are subject to change without notice.