Model 1016H-2 Temperature/Humidity Chamber Specifications
With optional GN2 Purge or Dry Air Purge: 5% to 95%
 (Dry bulb range of +10°C to +85°C)

NOTE: Ability to reach RH extremes may be limited by the humidity sensor accuracy. Low Dew Point conditions can only be achieved when starting with a clean, dry chamber.
| Cool Down Transition Time (empty, uncontrolled humidity mode)* | ||||||
| End Temperature | ||||||
| +23°C | 0°C | -20°C | -40°C | -55°C | -65°C | |
| Time from +23°C: °C/minute: | __ | 4 min 5.8 °C/min | 10 min 4.3°C/min | 18 min 3.5°C/min | 27 min 2.9°C/min | 35 min 2.7°C/min | 
| Time from +85°C: °C/minute: | 13 min 4.8°C/min | 18 min 4.3°C/min | 28 min 3.4°C/min | 38 min 3.3°C/min | 48 min 2.9°C/min | 60 min 2.5°C/min | 
| Time from +150°C: °C/minute: | 27 min 4.7°C/min | 34 min 4.4°C/min | 44 min 3.9°C/min | 54 min 3.5°C/min | 63 min 3.3°C/min | 73 min 2.9°C/min | 
| Heat Up Transition Time (empty, uncontrolled humidity mode)* | ||||||
| -40°C to +85 °C @ 208V input | 28 minutes (4.5°C/minute) | |||||
| *Note: Transition times are measured after a 2 hour soak at the respective start temperature with an empty chamber, as indicated on the temperature controller, 23°C ambient. Measured with setpoint beyond the end temperatures with an instantaneous step change in Set Point without a controlled ramp-rate. Does not include the effect of proportional band when approaching setpoint. Performance is reduced by 17% with 50Hz input power. | ||||||
| Live Load Capacity (uncontrolled humidity mode) | ||||||
| +23°C | 0°C | -20°C | -40°C | -55°C | -65°C | |
| 2,250 Watts | 1,890 Watts | 1,475 Watts | 1,100 Watts | 800 Watts | 540 Watts | |
| Refrigeration and Heating System | ||
| High Stage Refrigerant | R-404A (Dupont HP-62) | |
| Low Stage Refrigerant | R-508B (Dupont SUVA-95) | |
| Compressors | 2 HP x 2 HP Tecumseh compressors in a cascade configuration | |
| Condenser | Air Cooled | |
| Heat of Rejection | 17,485 BTUH (maximum rated chamber load at maximum cooling rate from high temperature soak) | |
| Heater Power | 2,820 Watts @ 208 V input; 3750 Watts @ 240 V input | |
| Air Flow | 830 cfm | |
| Instrumentation | ||
| Temp/Humidity Controller | Watlow F4T Touch Screen Controller with RS-232, Ethernet interface, 4.3" color graphic touch screen. OR... Watlow F4 Controller with RS-232 interface, LED readout of temperature, LCD display of other parameters. | |
| Limit Controller | Independent high and low temperature limits. Triggers an audible alarm and shuts down the chamber. Relay contacts provide a safety power interlock for test sample. | |
| Chart Recorder | (Optional) Honeywell DR4300 Series. Two pen, 10" circular chart. Mounts in lower front door. |  | 
| Input Power Requirements | |||
| Phase | Maximum Current Draw | Minimum Service | |
| 230 V ±10%, 60 Hz | 1 PH | 36 A | 45 A | 
| 3 PH | 24 A | 30 A | |
| 208 V -5/+10%, 60 Hz | 1 PH | 33 A | 45 A | 
| 3 PH | 22 A | 30 A | |
| Input may be configured for single or three phase in the field by changing jumpers. Three phase load is semi-balanced. Call for other voltages or 50 Hz operation. Performance is reduced by 17% at 50 Hz. Customer power source must be hard-wired to the chamber by a qualified electrician. Power cord and plug is not included. | |||
| Physical Characteristics and Safety | |
| Inside Dimensions | 30" W x 30" H x 30" D (15.6 cubic feet) 762 mm W x 762 mm H x 762 mm D (442 liters) | 
| Outside Dimensions | 38" W x 78.5" H x 56" D (nominal) 914 mm W x 1994 mm H x 1422 mm D Door latch adds 3" to width on right side. | 
| Minimum Installed Clearance | 18" from the left and right side 24" from the rear | 
| Window Viewing Area | 18" W x 12" H | 
| Access Ports | 4" Port on left and right side (two total). Supplied with foam plugs. | 
| Weight | Chamber Weight: 1,130 pounds Shipping Weight: 1,280 pounds | 
| Sound Level | 62 dBA in cooling mode (A-weighted, measured 36" from the front surface, 63" from the floor, in a free-standing environment) | 
NOTE: Performance is typical and based on operation at 23°C (73°F) ambient and nominal input voltage. Designed for use in a normal conditioned laboratory. Operation at higher ambient temperatures may result in decreased cooling performance. Additional ports and shelves will also affect performance. Operation above 30°C (85°F) or below 16°C (60°F) ambient is not recommended.
Due to continuous product improvements, specifications subject to change without notice.