Model 1016S Temperature Chamber Specifications


Temperature Range
-35°C to +175°C
Control Tolerance
±0.5°C (Short-term variations measured at the control sensor after stabilization)
Uniformity
±1.0°C (Variations throughout the chamber after stabilization, up to +8°C)
 
Cool Down Transition Time (empty)*
End Temperature
+23°C0°C-10°C-20°C-30°C-35°C
Time from +23°C:
°C/minute:
__1 min
23°C/min
2 min
16.5°C/min
4 min
10.8°C/min
8 min
6.6°C/min
Ultimate
Time from +85°C:
°C/minute:
4 min
15.5°C/min
7 min
12°C/min
9 min
10.6°C/min
12 min
8.8°C/min
17 min
6.8°C/min
Ultimate
Heat Up Transition Time (empty)*
-30°C to +85 °C13 minutes (8.8°C/minute)

*Note: Transition times are measured after a 2 hour soak at the respective start temperature with an empty chamber, as indicated on the temperature controller, 23°C ambient. Measured with setpoint beyond the end temperatures with an instantaneous step change in Set Point without a controlled ramp-rate. Does not include the effect of proportional band when approaching setpoint. Performance is reduced by 17% with 50 Hz input power.


 
Live Load Capacity
+23°C0°C-10°C-20°C-30°C
4,000 Watts2,800 Watts2,300 Watts1,800 Watts1,550 Watts

 
Refrigeration and Heating System
RefrigerantR-404A (Dupont HP-62)
Compressors3.5 HP Copeland scroll compressor. More about Scroll Compressors >>
CondenserAir Cooled
Heat of Rejection30,200 BTUH (maximum rated chamber load at maximum cooling rate from high temperature soak)
Heater Power4,200 Watts @ 208 V input
Air Flow 830 cfm
Instrumentation
Temperature ControllerWatlow F4T Touch Screen Controller with RS-232, Ethernet interface, 4.3" color graphic touch screen. OR...
Watlow F4 Controller with RS-232 interface, LED readout of temperature, LCD display of other parameters.
Limit ControllerIndependent high and low temperature limits. Triggers an audible alarm and shuts down the chamber. Relay contacts provide a safety power interlock for test sample.
Chart Recorder(Optional) Honeywell DR4300 Series. One pen, 10" circular chart. Mounts in lower front door.

 
Input Power Requirements
230 V ±10%, 60 Hz, 3 PhaseMax Current Draw 31 A; Recommended Service 40 A
208 V -5/+10%, 60 Hz, 3 PhaseMax Current Draw 26 A; Recommended Service 35 A

Input may be configured for 230 V or 208 V in the field by changing jumpers. Three phase load is balanced. Call for other voltages or 50Hz operation. Performance is reduced by 17% with 50Hz input power.


Customer power source must be hard-wired to the chamber by a qualified electrician. Power cord and plug is not included.


Physical Characteristics and Safety
Inside Dimensions30" W x 30" H x 30" D (15.6 cubic feet)
762 mm W x 762 mm H x 762 mm D (442 liters)
Outside Dimensions38" W x 78.5" H x 56" D (nominal)
914 mm W x 1994 mm H x 1422 mm D
Door latch adds 3" to width on right side.
Minimum Installed Clearance18" from the left and right side
24" from the rear
Window Viewing Area18" W x 12" H
Access Ports4" Port on left and right side (two total). Supplied with foam plugs.
WeightChamber Weight: 1,080 pounds
Shipping Weight: 1,220 pounds
Sound Level 67 dBA in cooling mode (A-weighted, measured 36" from the front surface, 63" from the floor, in a free-standing environment)

NOTE: Performance is typical and based on operation at 23°C (73°F) ambient and nominal input voltage. Designed for use in a normal conditioned laboratory. Operation at higher ambient temperatures may result in decreased cooling performance. Additional ports and shelves will also affect performance. Operation above 30°C (85°F) or below 16°C (60°F) ambient is not recommended.


Due to continuous product improvements, specifications subject to change without notice.