Model 1027S Temperature Chamber Specifications


Temperature Range
-35°C to +175°C
Control Tolerance
±0.5°C (Short-term variations measured at the control sensor after stabilization)
Uniformity
±1.0°C (Variations throughout the chamber after stabilization)
 
Cool Down Transition Time (empty)*
Start TempEnd Temp
+23°C0°C-10°C-20°C-30°C-35°C
+150°C9.0 min14 min17 min20 min26 minUltimate
+85°C4.0 min7.5 min10 min13 min19 min
+23°C-----1 min2 min4 min9 min

Cool Down Transition Time (with 80 lb. aluminum load)*
Start TempEnd Temp
+23°C0°C-10°C-20°C-30°C-35°C
+150°C20.0 min28 min33 min38 min46 minUltimate
+85°C8.0 min16 min20 min25 min33 min

Heat Up Transition Time (empty)*
Start TempEnd Temp
+23°C+50°C+85°C+125°C+150°C+175°C
+23°C-----1.5 min7 min14 min20 min25 min
0°C1.5 min3.5 min13 min20 min23 min31 min
-10°C2.2 min4.2 min14 min22 min25 min33 min
-35°C5.5 min10 min16 min23 min29 min34 min

Heat Up Transition Time (with 80 lb. aluminum load)*
Start TempEnd Temp
+23°C+50°C+85°C+125°C+150°C+175°C
0°C5 min13 min23 min36 min45 min55 min
-35°C11 min19 min29 min42 min51 min61 min

Rate Of Change
To calculate rate of change for a particular condition, take the difference between the Start Temp and End Temp and divide by the Transition Time.
Cool Down Example (empty): From +85°C to 0°C = 85°C / 7.5 min = 11.3°C/min.
Cool Down Example (with 80 lb. load): From +85°C to 0°C = 85°C / 16 min = 5.3°C/min.
Heat Up Example (empty): From 0°C to +85°C = 85°C / 13 min = 6.5°C/min.
Heat Up Example (with 80 lb. load): From 0°C to +85°C = 85°C / 23 min = 3.7°C/min.


*Note: Transition times are measured after a 2 hour soak at the respective start temperature with an empty chamber, as indicated on the temperature controller, 23°C ambient. Measured with setpoint beyond the start and end temperatures. Does not include the effect of proportional band when approaching setpoint. Performance is reduced by 17% with 50 Hz input power.


Live Load Capacity


+23°C0°C-10°C-20°C-30°C
4,000 Watts2,800 Watts2,300 Watts1,800 Watts1,250 Watts

Refrigeration and Heating System

RefrigerantR-404A (Dupont HP-62)
Compressors3.5 HP Copeland scroll compressor. More about Scroll Compressors >>
CondenserAir Cooled
Heat of Rejection30,200 BTUH (maximum rated chamber load at maximum cooling rate from high temperature soak)
Heater Power4,000 Watts @ 208 V input
Air Flow 830 cfm

Instrumentation

Temperature ControllerWatlow F4T Touch Screen Controller with RS-232, Ethernet interface, 4.3" color graphic touch screen. OR...
Watlow F4 Controller with RS-232 interface, LED readout of temperature, LCD display of other parameters.
Limit ControllerIndependent high and low temperature limits. Triggers an audible alarm and shuts down the chamber. Relay contacts provide a safety power interlock for test sample.
Chart Recorder(Optional) Honeywell DR4300 Series. One pen, 10" circular chart. Mounts in lower front door.

Input Power Requirements

230 V ±10%, 60 Hz, 3 PhaseMax Current Draw 31 A; Recommended Service 40 A
208 V -5/+10%, 60 Hz, 3 PhaseMax Current Draw 26 A; Recommended Service 35 A

Input may be configured for 230 V or 208 V in the field by changing jumpers. Three phase load is balanced. Call for other voltages or 50 Hz operation.

Customer power source must be hard-wired to the chamber by a qualified electrician. Power cord and plug is not included.


Physical Characteristics and Safety

Inside Dimensions40" W x 32" H x 36.5" D (27 cubic feet)
1016 mm W x 812.8 mm H x 927 mm D (764 liters)
Outside Dimensions49" W x 73.25" H x 63" D (nominal)
1244 mm W x 1860 mm H x 1600 mm D
Door latch adds 3" to width on right side. Circulator motor and housing adds 6" to height - may be removed for move-in.
Minimum Installed Clearance18" from the left and right side
24" from the rear
Window Viewing Area16" W x 13" H
Access Ports4" Port on left and right side (two total)
Supplied with foam plugs
WeightChamber Weight: 1,410 pounds
Shipping Weight: 1,595 pounds

NOTE: Performance is typical and based on operation at 23°C (73°F) ambient and nominal input voltage. Designed for use in a normal conditioned laboratory. Operation at higher ambient temperatures may result in decreased cooling performance. Additional ports and shelves will also affect performance. Operation above 30°C (85°F) or below 16°C (60°F) ambient is not recommended. Performance is reduced by 17% with 50 Hz input power.


Due to continuous product improvements, specifications subject to change without notice.