OMNIA II-01 Internal Scanner Option
The Scanner option provides 8 High-Voltage/Continuity channels and 8 Ground Bond/Continuity channels on the rear panel. The High Voltage/Continuity channels can be set to a High or Low level, giving the capability to test from one channel to another channel or from any channel to a common Low or Return 139 point. The channels can be connected in parallel if desired but there is only one leakage current measurement for all channels.
The Ground Bond/Continuity channels can only be used to test from one channel to the common Return. Only one channel may be used for each Ground Bond or Continuity test. The Ground Bond or Continuity test will be performed between the output channel selected and the Ground Bond/Continuity common terminal. Refer to the OMNIA II Series pdf manual for a full test instructions for this option.
Ground Bond/Continuity channel Resistance Deviation:
- Deviation (Relay Contact): Test Current
- 6 megohm maximum: > or equal 10A