Temperature Range
-73°C to +175°C
Control Tolerance
±0.5°C (±0.2°C Typical) (Short-term variations measured at the control sensor after stabilization)
Uniformity
±1.0°C (Variations throughout the chamber after stabilization)
Heat Up Rate of Change*
5°C/minute typical
Cool Down Rate of Change*
2.8°C/minute from +85°C to -40°C (standard 60 Hz models)
2.1°C/minute from +85°C to -40°C (115R-EX 50 Hz export models)
Cool Down Transition Times*
Start TempEnd Temp
+23°C0°C-40°C-55°C-65°C-73°C
+23°C-----5 min20 min30 min40 minUltimate
+85°C16 min25 min45 min57 min69 minUltimate

Rate Of Change
To calculate rate of change for a particular condition, take the difference between the Start Temp and End Temp and divide by the Transition Time.
Cool Down Example: From +85°C to -40°C = 125°C / 45 min = 2.8°C/min.

*Note: Transition times are measured after a 1 hour soak at the respective start temperature with an empty chamber, as indicated on the temperature controller, 23°C ambient. Measured with setpoint beyond the start and end temperatures. Does not include the effect of proportional band when approaching setpoint. Measured at nominal line voltage of 120V, 60 Hz. Performance is reduced on 115R-EX 50 Hz export versions.
 
Live Load Capacity
+23°C0°C-40°C-55°C-65°C
300 Watts255 Watts160 Watts110 Watts80 Watts
 
Refrigeration and Heating System
High Stage RefrigerantR-404A (Dupont HP-62)
Low Stage RefrigerantR-508B (Dupont SUVA-95)
Compressors1/2 HP x 1/3 HP Tecumseh hermetic compressors in a cascade configuration.
CondenserAir Cooled
Heat of Rejection4,000 BTUH (maximum rated chamber load at maximum cooling rate from high temperature soak)
Heater Power600 Watts
Air Flow85 CFM
 
Instrumentation
Temperature ControllerWatlow F4T Touch Screen Controller with RS-232, Ethernet interface, 4.3" color graphic touch screen. OR...
Watlow F4 Controller with RS-232 interface, LED readout of temperature, LCD display of other parameters.
Limit ControllerIndependent high and low temperature limits. Triggers an audible alarm and shuts down the chamber.
 
Power Requirements
Input Voltage

120 V nominal (110 to 126 VAC), 60 Hz, 1 PH
Max Current Draw 15 A, Recommended Minimum Service 20 A

Model 115-EX export version available with 230V, 50Hz, 8A input. Performance is reduced by 17% with 50 Hz input power.

Power Cord
and Plug

NEMA 5-20P Plug, NEMA 5-20P6' Power cord supplied with a molded NEMA 5-20P plug.

Plugs into a standard NEMA 5-20R receptacle.
Use of an extension cord is not recommended.

Model 115R-EX export version is supplied with power cord for the destination country.

Physical Characteristics and Safety
Inside Dimensions16" W x 12" H x 14" D (1.55 Cubic feet)
406 mm W x 305 mm H x 353 mm D (44 liters)
Outside Dimensions

Benchtop Model 115R

24" W x 31.6" H x 50" D (nominal)
610 mm W x 803 mm H x 1270 mm D

Minimum Installed Clearance12" from the right and left side
6" from the rear
Window Viewing Area8" W x 6" H
Access Ports3" (2.83" inside diameter) supplied on all Model 115R chambers. Port on left and right side (two total)
Supplied with silicone foam port plugs
Weight

Benchtop Model 115R

Chamber Weight: 300 pounds

Shipping Weight: 375 pounds

Sound Level52 dBA in cooling mode (A-weighted, measured 36" from the front surface, 63" from the floor, in a free-standing environment)
 

NOTE: Performance is typical and based on operation at 23°C (73°F) ambient and nominal input voltage. Designed for use in a normal conditioned laboratory. Operation at higher ambient temperatures may result in decreased cooling performance. Additional ports and shelves will also affect performance. Operation above 30°C (85°F) or below 16°C (60°F) ambient is not recommended.

Due to continuous product development, specifications are subject to change without notice.