Model 1007C Temperature Chamber Specifications


Temperature Range
-73°C to +175°C

Control Tolerance
±0.5°C (±0.2°C Typical) (Short-term variations measured at the control sensor after stabilization)

Uniformity
±1°C (Variations throughout the chamber after stabilization)
 
Cool Down Transition Time*
Start TempEnd Temp
+23°C0°C-40°C-55°C-65°C-73°C
+23°C-----4 min18 min25 min33 minUltimate
+50°C5 min11 min27 min36 min45 minUltimate
+85°C13 min20 min37 min47 min55 minUltimate
+150°C32 min39 min58 min65 min76 minUltimate
Heat Up Transition Time*
Start TempEnd Temp
+23°C+50°C+85°C+125°C+150°C+175°C
+23°C-----1.5 min7 min14 min20 min25 min
0°C1.5 min3.5 min13 min20 min23 min31 min
-40°C6 min11 min17 min24 min30 min35 min
-55°C8 min13 min19 min26 min32 min37 min
-65°C10 min14 min21 min28 min34 min39 min

Rate Of Change
To calculate rate of change for a particular condition, take the difference between the Start Temp and End Temp and divide by the Transition Time.
Cool Down Example: From +50°C to -40°C = 90°C / 27 min = 3.33°C/min.
Heat Up Example: From -40°C to +50°C = 90°C / 11 min = 8.18°C/min.


*Note: Transition times are measured after a 2 hour soak at the respective start temperature with an empty chamber, as indicated on the temperature controller, 23°C ambient. Measured with setpoint beyond the start and end temperatures. Does not include the effect of proportional band when approaching setpoint. Performance is reduced by 17% with 50 Hz input power.

 
Live Load Capacity
+23°C0°C-40°C-55°C-65°C
1,000 Watts800 Watts500 Watts400 Watts300 Watts
 
Refrigeration and Heating System
High Stage RefrigerantR-404A (Dupont HP-62)
Low Stage RefrigerantR-508B (Dupont SUVA-95)
Compressors1.5 HP x 1.5 HP Tecumseh hermetic compressors in a cascade configuration.
CondenserAir Cooled
Heat of Rejection14,800 BTUH (maximum rated chamber load at maximum cooling rate from high temperature soak)
Heater Power2,000 Watts
Air Flow 450 cfm
 
Instrumentation
Temperature ControllerWatlow F4T Touch Screen Controller with RS-232, Ethernet interface, 4.3" color graphic touch screen. OR...
Watlow F4 Controller with RS-232 interface, LED readout of temperature, LCD display of other parameters.
Limit ControllerIndependent high and low temperature limits. Triggers an audible alarm and shuts down the chamber. Relay contacts provide a safety power interlock for test sample.
Chart Recorder (Optional) Honeywell DR4300 Series. One pen, 10" circular chart. Mounts in lower front door.Honeywell Chart Recorder
 
Input Power Requirements
PhaseCurrent DrawMinimum Service
230 V ±10%, 60 Hz1 PH25 A30 A
3 PH19 A25 A
208 V -5/+10%, 60 Hz1 PH28 A35 A
3 PH21 A30 A

Input may be configured for single or three phase in the field by changing jumpers. Three phase load is semi-balanced. Call for 50 Hz operation. Performance is reduced by 17% with 50 Hz input power.

Customer power source must be hard-wired to the chamber by a qualified electrician. Power cord and plug is not included.

 
Physical Characteristics and Safety
Inside Dimensions24" W x 21" H x 24" D (7 cubic feet)
609 mm W x 533 mm H x 609 mm D (198 liters)
Outside Dimensions*33" W x 67.75" H x 54.25" D (nominal)
838 mm W x 1721 mm H x 1378 mm D
* Door latch adds 3" to width on right side (may be removed to permit move-in through a 36" doorway). Circulator motor and housing adds 6" to height.
Minimum Installed Clearance18" from the left and right side
24" from the rear
Window Viewing Area13.375" W x 9" H
Access Ports4" Port on left and right side (two total)
Supplied with foam plugs
WeightChamber Weight: 850 pounds
Shipping Weight: 1,011 pounds
Sound Level62 dBA in cooling mode (A-weighted, measured 36" from the front or side surface, 63" from the floor, in a free-standing environment)

NOTE: Performance is typical and based on operation at 23°C (73°F) ambient and nominal input voltage. Designed for use in a normal conditioned laboratory. Operation at higher ambient temperatures may result in decreased cooling performance. Additional ports and shelves will also affect performance. Operation above 30°C (85°F) or below 16°C (60°F) ambient is not recommended.